At the start of the cycle the chamber is 100% air and the cycle
is designed to remove as much as possible. This is done by using a vacuum pump
and pulsing steam into the chamber. Additional air can come into the chamber via leaks, as non
condensable gases in the steam supply or be generated from load items.
An 
air detector is used to detect the
presence of air in a porous load cycle. This is crucial because air pockets can
prevent sterilization. There are two main types of air detectors:
- Temperature-Based Air Detector: This
     method creates a "worst-case" scenario location where air is
     likely to collect. The temperature is then measured in this area; if air
     is present, it will show a colder temperature compared to the
     surrounding steam.
 
 
- Pressure-Based Air Detector: This
     method works by collapsing a steam sample and measuring the pressure. The
     more non-condensable gases (such as air) present in the steam, the
     higher the pressure will be.
 
 
The air detector compares the results against
pre-defined setpoints. If the setpoints are met, the cycle continues as normal.
If not, the cycle will be aborted, and a cycle failure will occur.
For detailed information on air detectors and
their setup, refer to 
ISO EN-285. Or watch our 
Free webinar 'Introduction Air Detectors'.
You can also learn more about air
detectors and their configuration in our 
Advanced Autoclave Validation
training course.