We use Ellab Tracksense Pro dataloggers with response times of T-90 of ≤5 Seconds. At Thermal Compliance we have many different sensors all with a response time T-90 between 0.8-4.3 seconds.
In addition to the benefits of response time, using dataloggers allows us to:
•Reduce Validation time by up to 50%. •Work around production requirements. •Validation work can be performed offline, freeing up access to process, by performing: o Offline Pre and Post Calibration. o Offline Load Preparation. o Offline Sensor and/or Biological Indicator Placement. o Reduced Leak Rate Tests •Sky wireless system shows live data. •Faster data processing, with clear and easy to review reports. |
If you would like to know more about our validation services or the benefits of using wireless dataloggers please contact Chris Maughan
Section 23.3.3.1.1 EN285:2015 requires a sensor response time T-90 of ≤5 Seconds. We perform independent response time testing, if you would like to know more please visit.